Clinical Report: Fitting Irregular Corneas Secondary to Penetrating Keratoplasty
Overview
This case report evaluates the fitting of a protruding, tilted corneal graft using an oblate semiscleral contact lens. The integration of corneal tomography and AS-OCT facilitated lens fitting and monitoring of corneal health.
Background
Fitting irregular corneas after penetrating keratoplasty (PK) poses challenges due to complex topographies resulting from graft tilt or elevation. Traditional contact lens options often fail.
Data Highlights
| Parameter | Measurement |
|---|---|
| Initial Base Curve | 7.46 mm |
| Final Base Curve | 7.1 mm |
| Diameter | 15.0 mm |
| Power | –7.75 D |
| Nasal Clearance (Initial) | 192 µm |
| Nasal Clearance (14 weeks) | 148 µm |
Key Findings
- The ROSE K2 XL Oblate lens was effective for fitting a tilted graft post-PK.
- Initial fitting attempts with GP, soft toric, and hybrid lenses were unsuccessful due to mechanical strain.
- Corneal tomography and AS-OCT were crucial for assessing corneal elevation and lens clearance.
- Final lens fitting achieved stable visual acuity of 6/9+ (20/30+).
- The patient achieved 14 hours of daily wear with excellent comfort and no corneal staining.
Clinical Implications
The use of oblate semiscleral lenses can provide a fitting strategy for patients with irregular corneas following PK.
Conclusion
This case demonstrates the effectiveness of semiscleral lenses in managing complex corneal grafts.
Related Resources & Content
- Contact Lens Spectrum, 1997 -- Managing The Irregular Cornea With Contact Lenses
- Contact Lens Spectrum, 1998 -- Contact Lenses for the Irregular Cornea
- Contact Lens Spectrum, 2026 -- Using Specialty Soft Lenses for Irregular Cornea
- EyeWiki -- Penetrating Keratoplasty
- PubMed -- Contact lenses for visual rehabilitation in post-keratoplasty eyes: A systematic review
- Contact Lens Spectrum — Management of Corneal Irregular Astigmatism
- Penetrating Keratoplasty - EyeWiki
- Contact lenses for visual rehabilitation in post-keratoplasty eyes: A systematic review - PubMed
- Corneal and Intraocular Pressure Responses to Scleral Lens Wear: A Meta-Analysis | Ophthalmic and Physiological Optics | Springer Nature Link
This content is an AI-generated, fully rewritten summary based on a published scholarly article. It does not reproduce the original text and is not a substitute for the original publication. Readers are encouraged to consult the source for full context, data, and methodology.